MPX Series Magnetostrictive Level Sensors Rev. B, 4/13
16
Automation Products Group, Inc.
APG...Providing tailored solutions for measurement applications
Modbus Register Lists
Input Registers (0x04):
Register Returned Data
30300 Raw Top Float Reading (in mm, unsigned)
30301 Raw Boom Float Reading (in mm, unsigned)
30302 Temperature Reading (in
0
C, signed)
30303-30304 Calculated Top Float Reading (in selected Units)
30305-30306 Calculated Boom Float Reading (in selected Units)
NOTE: the Calculated Readings will be returned without a decimal place. In
order to aain the true result, the Decimal Place seng must be taken into
account. Refer to the Decimal Place seng on page 19 for more informaon.
Holding Registers (0x03):
Register Funcon Value Range
40400 Device Address 1 to 255
40401 Units 1 to 3
40402 Applicaon Type 0-10
40403 Volume Units 0 to 6
40404 Decimal Place 0 to 10364 mm
40405 Max Distance 0 to 10364 mm
40406 Full Distance 0 to 10364 mm
40407 Empty Distance 0 to 10364 mm
40408 Sensivity 0 to 100
40409 Pulses 0 to 20
40410 Blanking 0 to 10364 mm
40411 Reserved n/a
40412 Averaging 0 to 100
40413 Filter Window 0 to 10364 mm
40414 Out of Range Samples 0 to 255